Go to content

Go to content | Navigation | Direct access | Connection

You are here : OVSQENPITPresentation of the Technical ResourcesOptical testing

Optical testing

Monochromator


The PIT has a monochromator dedicated to high spectral resolution optical tests: instrument calibration, detector and source characterisation. This equipment, which can be used independently, could also be coupled with the thermal-optical vacuum vessel in order to provide a means of calibration which can be used in a thermal environment.
 


Photo: PIT

Characteristics of the Monochromator

  • Mc Pherson model type 225 with 1m focal length under vacuum
  • Vacuum: P <10-5 mbar
  • 4 networks available to cover the spectral range 30 to 1200 nm
  • Adjustable slits from 5 µm to 2 mm
  • Focal plane 50mm wide
  • Resolutions: 30 to 1400 nm
  • Accuracy: +/- 0.005 nm